Our Products
   
By Product Type, Brand or Keyword
Services

Featured Products & Specials
Parts, Accessories, Rental, Repair
Our Quality testing process!
Request a Quote / More Information
Customer Survey
Operations Manuals - Download

Request an RMA

Ardus Medical
 CADD Pumps I CADD Legacy Plus Pump, Ambulatory Pump

CADD Legacy Plus Ambulatory Pump

Cadd Legacy Plus

Cadd Legacy Plus Ambulatory Pump

Save up to 80% on pre-owned Cadd Legacy Plus ambulatory pumps.
   Request A QUOTE!

Product Specifications

  • Pump Resolution:
  • 0.050 ml per pump stroke nominal
  • Dimensions:
  • 1.6" D x 3.8" W x 4.4" H excluding cassette
  • Weight:
  • 13.8 oz including 2 AA batteries and empty 100 ml reservoir
  • Power Source:
  • 2 AA alkaline batteries or AC Adapter
  • Maximum Infusion Pressure:
  • 40.0 psi
  • System Delivery Accuracy:
  • ±6% nominal
  • Maximum Time to Occlusion Alarm:
  • 2.0 hours
  • Delivery Rate During Priming:
  • Approximately 180 ml/hr

Continuous Delivery Mode Specifications

  • Reservoir Volume:
  • 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
  • Continuous Rate:
  • 0.1 - 125 ml/hr; programmable in 0.1 ml/hr increments
  • Given:
  • 0-99999.95 in 0.05 ml increments

Intermittent Delivery Mode Specifications

  • Reservoir Volume:
  • 1-9999 or Not in Use; programmable in 1 ml increments, displayed in 0.1 ml increments
  • Dose Volume:
  • 0.1 - 1000 ml programmable in 0.1 ml increments
  • Dose Duration:
  • 1 min to 24 hrs in the following increments
  • - 1 minute for values between 1 min and 10 min
  • - 5 minutes for values above 10 min
  • Dose Cycle:
  • 10 min to 96 hrs in 5 min increments
  • KVO Rate:
  • 0 to 125 ml/hr in 0.1 ml/hr increments
  • Dose Starts In:
  • Immediate or 1 min to 96 hrs in the following increments
  • - 1 min for values between 0 min and 10 min
  • - 5 min for values between 10 min and 96 hrs
  • Given:
  • 0-99999.95 in 0.05 ml increments
  • Dose Remaining:
  • 1 minute increments
  • Pump Alarms:
  • Low battery power
  • depleted battery power
  • battery dislodged
  • pump stopped
  • pump fault
  • low reservoir volume
  • high delivery pressure
  • air in line
  • disposable not attached when run attempted
  • motor locked
  • upstream occlusion
  • reservoir volume empty
  • program incomplete
  • key stuck
  • disposable detached

For service, parts, PM or rental see our Services section


      Slothworks
Copyright © 2010 All Rights Reserved. Ardus Medical, Inc. | 1-855-59-Ardus | info@ardusmedical.com | Site Map