Dimensions:
Approximately 8.25H x 7.5W x 8.75D inches (including pole clamp)
Weight:
Approximately 7.75 lbs (with battery)
Casing:
High-impact plastic
Power Requirements:
100-130 Vac, 50 to 60 Hz, less than 35 W
Power Cord:
Hospital-grade AC (mains) cord. 10 feet long, with transparent plug
Fuses:
0.5 A, 250 V. slo-blo
Battery:
Sealed lead-acid rechargeable 8 V battery, internal to infusion system. Accessible for ease of field replacement, with color-coded leads and polarized connector.
Battery Operation:
A fully charged new battery will provide eight hours of operation at 125 ml/hr, or 1000 ml total volume delivered, whichever occurs first
Recharge:
The battery charges whenever the infusion system is connected to AC (mains) power. If the infusion system is operating at 125 ml/hr, a full recharge takes approximately 16 hours. If the infusion system is turned to OFF/CHARGE, recharge takes approximately eight hours.
Delivery Rate Range
Micro Mode:
.1 to 99.9 ml/hr
Macro Mode:
1 to 999 ml/hr (in 1 ml increments)
KVO:
1 ml/hr
Dose Limit Range:
1 to 9999 ml/hr (in 1 ml increments)
Occlusion Range:
10 psig (+5, -2 psig)