Product Specifications
- Dimensions:
- 6"H x 4"W x 1.5"D
- Power Sources:
- Wall Plug-in AC Power Supply, 120 VAC, 60 Hz, 0.06A
- Battery:
- Four AA Alkaline Batteries
- Battery Pack:
- Rechargeable using AC Power Supply
- Power Capacity:
- Four AA Alkaline Batteries or Battery Pack, at least 3400 mL at rates above 50 mL/hr and at least four days at rates below 10 mL/hr
- Pump Mechanism:
- Wall Plug-in AC Power Supply, 120 VAC, 60 Hz, 0.06A
- Memory Protection:
- Current Program and 512 event history log protected by internal battery-backed memory for at least one year after power is removed
- System Accuracy:
- ±5% nominal
TPN Program Specifications
- Total Infusion Time:
- 1 min - 24 hours
- Taper Up/Down Times:
- 1 min - 3 hours
- KVO Rate:
- 1.0 ml/hr - 5.0 ml/hr
- KVO Time:
- 1 min - 24 hours
- Container Size:
- TPN plus KVO Volume - 9999 ml
Pain Management Specifications
- Concentration:
- 0.1 mg/ml - 50 mg/ml, or 1 µg/ml - 1000 µg/ml
- Delivery Rate:
- 0.0 ml/hr continuous plus bolus or 0.1 ml/hr continuous only - 5.0 ml/hr subq or 25.0
- ml/hr epidural, or intravenous
- 0.0 mg/hr - 1250 mg/hr, or 1 µg/hr - 25000 µg/hr
- Loading and Bolus Doses:
- 0.0 ml/hr continuous plus bolus or 0.1 ml/hr continuous only - 5.0 ml/hr subq or 25.0 ml/hr epidural, or intravenous
- ml/hr epidural, or intravenous
- 0.0 mg - 1250 mg, or 1 µg - 25000 µg
- Loading & Bolus Doses Rate:
- approximately 125 ml/hr
- Bolus Lockout:
- 5 min - 999 min
- Boluses Per Hour Limit:
- 1 per hour - 12 per hour
- Container Size:
- 0.1 ml - 9999.9 ml
- 0.1 mg - 9999.9 mg, or 1 µg - 999999 µg
Intermittent Delivery Specifications
- Concentration:
- 0.1 mg/ml - 1000.0 mg/ml, or 1 µg/ml - 1000 µg/ml
- Dose Amount:
- 0.1 ml - 9600 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 999900 µg
- Infusion Time Per Dose:
- 1 min - 24 hours
- Dose Frequency:
- 1 min - 96 hours
- Number of Doses in Container:
- 1 dose - 999 doses
- KVO Rate:
- 0.1 ml/hr - 5.0 ml/hr
- 0.1 mg/hr - 5000 mg/hr, or 1 µg/hr - 5000 µg/hr
- Container Size:
- 0.1 ml - 9999.9 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 999900 µg
Variable Time Delivery Specifications
- Concentration:
- 0.1 mg/ml - 1000 mg/ml, or 1 µg/ml - 1000 µg/ml
- Base Start/Stop Times:
- 12:00AM - 11:59PM
- Base Rate:
- 0.1 ml/hr - 400 ml/hr
- 0.1 mg/hr - 9900 mg/hr, or 1 µg/hr - 400000 µg/hr
- Daily Phase Amount:
- 0.1 ml - 9600 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 999900 µg
- Phase Start/Stop Times:
- 12:00AM - 11:59PM
- Phase Percentage:
- 0.1% - 100%
- Phase Rate:
- 0.1 ml/hr - 400 ml/hr
- 0.1 mg/hr - 9900 mg/hr, or 1 µg/hr - 400000 µg/hr
- Phase Dose:
- 0.1 ml - 9600 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 999900 µg
- Bolus Dose:
- 0.1 ml - 25 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 25000 µg
- Bolus Dose Delivery Rate:
- Approximately 400 ml/hr
- Bolus Lockout:
- 5 min - 999 min
- KVO Rate:
- 0.1 ml/hr - 5 ml/hr
- 0.1 mg/hr - 5000 mg/hr, or 1 µg/hr - 5000 µg/hr
- Container Size:
- 0.1 ml - 9999.9 ml
- 0.1 mg - 9900.0 mg, or 1 µg - 999900 µg
Continuous Delivery Specifications
- Concentration:
- 0.1 mg/ml - 1000 mg/ml, or 1 µg/ml - 1000 µg/ml
- Rate:
- 0.1 ml/hr - 400 ml/hr
- 0.1 mg/hr - 9900 mg/hr, or 1 µg/hr - 400000 µg/hr
- Container Size:
- 0.1 ml - 9999.9 ml
- 0.1 mg - 9900 mg, or 1µg - 999900 µg
For service, parts, PM or rental see our Services section |